Model Number LTV 1200
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to vyaire medical that the laptop ventilator sprintpack battery was no longer charging, and that the vent inoperable light was on, the charge status light was blinking orange, and no lights came on when the external battery was tested.Furthermore, there was no patient associated with the reported event.
 
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).The customer confirmed that the device will not be returned for further evaluation; therefore a definitive root cause could not be determined and corrective or preventive action is not indicated at this time.However, the customer stated that the device issue had been resolved and that a new battery would be shipped.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1200 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key15481440
Report Number2021710-2022-16651
Device Sequence Number6839835
Product Code CBK
UDI-Device Identifier10190752115279
UDI-Public(01)10190752115279(11)E07682
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K060647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLTV 1200
Device Catalogue Number18888-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2022
Initial Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1