| Model Number |
7K62-35 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a falsely decreased architect tsh result on a patient.Results provided: sid 46 = 0.066 uiu/ml, x5 dilution = 0.2717 uiu/ml (reference range: 0.35-4.94 uiu/ml).Other platform is in the normal range.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause architect tsh reagent lot 35087ud00 identified normal complaint activity and there are no trends for the product related to patient results.No customer returns were available for evaluation.Inhouse testing determined that the accuracy performance is not negatively impacted.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lot.This review did not identify any non-conformances, potential non-conformances or deviations.A review of labeling concluded that the issue is sufficiently addressed.Based on our investigation no systemic issue or product deficiency of the architect tsh reagent lot 35087ud00 was identified.
|
| |
|