Model Number D138502
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2022
Event Type  malfunction  
Event Description
It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath - medium.The hemostatic valve broke when catheter was removed.It was reported that the hemostatic valve on the vizigo sheath broke when trying to insert the catheter into the sheath.The sheath was replaced, and the issue resolved.While the physician was pulling the catheter out the sheath, the bwi company representative believes that the hemostatic valve dislodged inside the hub.No photos unfortunately.Sheath was inside the patient.No air entered the patient.No patient consequences were reported.Hemostatic valve separation is mdr-reportable.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
On 17-oct-2022, the bwi product analysis lab received the device for evaluation.It was reported that an unknown patient underwent an atrial fibrillation (afib) ablation procedure with a carto vizigo¿ 8.5f bi-directional guiding sheath - medium.The hemostatic valve broke when catheter was removed.Device evaluation details: visual analysis revealed that the hemostatic valve was dislodged inside of hub component.Microscopic examination of the hemostatic valve surface showed stress marks on the outer diameter.The damage observed could be related to the incorrect insertion of the dilator into the sheath causing the dislodgment of the valve; the stress marks and physical damage observed suggest that excessive force or manipulation was applied; however, this could not be conclusively determined.A device history record evaluation was performed for the finished device number 50000178 and no internal action related to the complaint was found during the review.The issue reported by the customer was confirmed.The odp (optimal device performance guide) contains the following caution: always insert a dilator straight into the center of the sheath¿s valve to prevent damage to the valve.Do not insert a dilator at an angle, as damage to the sheath valve may occur.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (IRVINE)
33 technology drive
irvine CA 92618
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-798
MDR Report Key15482625
MDR Text Key305626508
Report Number2029046-2022-02318
Device Sequence Number1
Product Code DYB
UDI-Device Identifier10846835016277
UDI-Public10846835016277
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170997
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberD138502
Device Catalogue NumberD138502
Device Lot Number50000178
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/06/2022
Initial Date FDA Received09/26/2022
Supplement Dates Manufacturer Received10/17/2022
01/22/2024
Supplement Dates FDA Received11/10/2022
01/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-