Based on the log file analysis the case could be reconstructed as follows: during downward movement of the piston the device detected that a negative airway pressure has been built-up to a certain level which is an indication for a fresh gas deficit.In such situation the device is designed to open an auxiliary air intake valve to compensate the fresh gas deficit by taking-in ambient air.It can be assumed that for the case in question this did not work or the air intake was not proceeded fast enough, resulting in a further increase of the negative airway pressure.The device consequently stopped piston movement temporarily and posted a "vent fail" alarm.If the negative pressure again falls below the threshold ventilation will resume automatically.As the paw sensor or auxiliary air valve potentially can contribute to the described symptom these components were tested on-site by a technician without finding any deviations from specification.During a simulated ventilation the issue could also not be duplicated.Additionally, the device was checked by the biomed over night whereby no alarms or issues were identified.Finally, the root cause for the reported symptom could not be determined.The unit was released for use without further probelem reported since then.
|