| Model Number |
02-012-35-2511 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Aseptic loosening (4883)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this 78 y/o female patient has had a total right knee revision due to femoral loosening.Patient was last known to be in stable condition following the event.Devices will not be returning, hospital disposed of them.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant device(s): 02-010-01-0225 - logic femoral ps cem left sz 2.5, 2887011; 02-012-45-2515 - lgc tibial fit tray cem sz 2.5f / 1.5t, 3706015; 200-02-32 - three peg patella 32mm, 3000676; 204-34-02 - fluted stem extension 25l x 14 mm 3726290.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the revision reported in was likely the result of patient-related conditions, an insufficient bond between the implant and the bone, or a combination of these possibilities, which led to aseptic (non-infected) femoral loosening.However, this cannot be confirmed as the devices were not returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d1, d4, d6a, h4, h6: clinical code, types of investigation, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6: investigation findings.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|