• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTR ANCH,PEEK SWI-VELOCK,3.5X 15.8MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTR ANCH,PEEK SWI-VELOCK,3.5X 15.8MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number SUTR ANCH,PEEK SWI-VELOCK,3.5X 15.8MM
Health Effect - Clinical Codes Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022, it was reported by an arthrex employee via email that an ar-2325pslc suture anchor broke during insertion.This was discovered during a case and broke inside the joint space.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation of the pictures of an ar-2325pslc attached to the complaint.It is inferred that the swivelock screw was damaged at the distal end at the tip.The probable cause of the event could not be determined from the information available and without device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH,PEEK SWI-VELOCK,3.5X 15.8MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15484168
Report Number1220246-2022-05523
Device Sequence Number7755604
Product Code MAI
UDI-Device Identifier00888867026971
UDI-Public00888867026971
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101823
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberSUTR ANCH,PEEK SWI-VELOCK,3.5X 15.8MM
Device Catalogue NumberAR-2325PSLC
Device Lot Number12583976
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date03/16/2023
Date Device Manufactured01/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-