TELEFLEX MEDICAL EZ-BLOCKER KIT; EZ BLOCKER Back to Search Results
Model Number IPN046435
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the: "ez blocker balloons were independently inflating and/or deflating on their own.".
 
Additional Manufacturer Narrative
Qn#(b)(4).
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed on a potential lot number reported by the customer and no relevant findings were identified.The manufacturing site reports that "based on communication with supplier of catheters, the reported defect was confirmed as "supplier related.Defect is probably caused by small cross-leak between balloons." it was also reported that the supplier opened a non-conformance in order to further investigate.
 
Event or Problem Description
It was reported that the: "ez blocker balloons were independently inflating and/or deflating on their own." no patient injury reported.The condition of the patient was listed as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ-BLOCKER KIT
Common Device Name
EZ BLOCKER
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15484450
Report Number3006425876-2022-00836
Device Sequence Number20866290
Product Code CBI
UDI-Device Identifier14026704589958
UDI-Public14026704589958
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K121462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN046435
Device Catalogue NumberMG-02770-002
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-