• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187644
Medical Device Problem Codes Delivered as Unsterile Product (1421); Unsealed Device Packaging (1444); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device 2 of 5.Complainant city: (b)(6).Complainant country: mexico.Based on the available information, this event is deemed to be a reportable malfunction.Per clinical team, based on photos, seal not damaged was not guaranteed.Hence, this case is being submitted for the same.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that the primary packaging was crushed.The customer did not wish to provide further information.The product was not used.The photographs depicting the issue were received from the complainant.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary photograph, video and/or physical sample evaluation: there are photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 1m00624 was manufactured on 06/dec/2021, in bopack line, with a total of (b)(4) market units (mkus).The complaint investigator performed a batch record review on 03/oct/2023, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material id 1704758 and manufacturing order (b)(4).The production process, in-process control, testing results and packaging of products was run according to the applicable process instruction (automated packaging bopack sterile and non-sterile products), review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 03/oct/2023, complaint investigator ran a query in database from 01/jan/2022 to 03/oct/2023 in order to verify the complaints reported for the lot number 1m00624 for the malfunction code ¿primary pack (sterile products) has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination (e.G.Water marks, stains), or dressing or loose material is trapped in packaging¿ and as result, no additional complaints were found.Historical nonconformance review: on 03/oct/2023, complaint investigator ran a query in database looking for any in process nonconformance / corrective action / preventive actions (capa) (s) associated to the malfunction code ¿primary pack (sterile products) has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination (e.G.Water marks, stains), or dressing or loose material is trapped in packaging¿ for the lot number 1m00624 and as result, no nonconformance / corrective action / preventive actions (capa) (s) for this malfunction code were generated during the manufacturing process of the referenced lot.Current quality controls: based on the process instruction (automated packaging bopack sterile and non-sterile products), the following tests are performed in the manufacturing lines, in order to identify this failure mode in our manufacturing process: tm - 002 ¿package seal integrity nonconformities- method 7: ¿ frequency: first piece, each 30 minutes, at the beginning of the shift.¿ sample quantity: 5 samples.¿ specification: visual - nlt 2 per cavity / manual peel.¿ acceptance criteria: accept = 0 | reject = 1.Defect rate analysis: there have only been 5 defective parts confirmed to date from a lot size of 12,500 products.This represents a defect rate of only 0.04% which is well within an appropriate acceptable quality level (aql) for primary pack (sterile products) has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination (e.G.Water marks, stains), or dressing or loose material is trapped in packaging issues which should be 0.25% based on our standard operating procedure (sop).In addition, all of the in-process testing on this lot did not find an adhesive issues, which confirms that the lot is unlikely to breach an acceptable quality level (aql) of 0.25.This issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted acceptable quality level (aql) level for this type of failure mode or defect.Conclusions: there are photographs associated with this case and in these, the reported defect can be seen and confirmed.The review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No nonconformity has been registered during the manufacturing process of the affected lot for the malfunction code ¿primary pack (sterile products) has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination (e.G.Water marks, stains), or dressing or loose material is trapped in packaging¿.No additional complaints were reported for lot affected related to the malfunction code ¿primary pack (sterile products) has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination (e.G.Water marks, stains), or dressing or loose material is trapped in packaging¿.Based on this, no negative trend was identified.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODERM
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key15484567
Report Number9618003-2022-05081
Device Sequence Number20136695
Product Code NAD
UDI-Device Identifier00768455174751
UDI-Public00768455174751
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number187644
Device Lot Number1M00624
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/31/2022
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-