SYNTHES GMBH 2.5MM DRILL BIT/QC/GOLD/110MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 310.250 |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional product codes: hsz, gff, hwe.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, during surgery, the screw that was previously assembled with stardrive recess was damaged.The drill bit was also dull.These events did not adversely affect the patient or the procedure.This report involves one 2.5mm drill bit/qc/gold/110mm.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: part: 310.250.Lot no: l272797.Release to warehouse date: 26 january 2017.Manufacturing site: werk bettlach.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo revealed that there is no damage or defects on the surface of the drill bit ø2.5 l110/85 2flute f/quick co.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed for drill bit ø2.5 l110/85 2flute f/quick co, p/n: 310.250.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 1 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the edges of the flutes of drill bit ø2.5 l110/85 2flute f/quick co appear to be softened and dull.A dimensional inspection for the drill bit ø2.5 l110/85 2flute f/quick co was unable to be performed due to post manufacturing damage.A functional test was unable to be performed due to the nature of its functionality.Nevertheless, as the edges of the flutes appear to be softened and worn it is reasonable that the dullness is attributed to this condition.The complaint condition was able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit ø2.5 l110/85 2flute f/quick co would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a.Device history lot.Part: 310.250.Lot: 5l22644.Manufacturing site: bettlach.Release to warehouse date: 26 july 2019.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10: concomitant device added.
|
|
Event Description
|
This is report 1 of 3 for (b)(4).
|
|