| Model Number |
382400 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Upon inspection of the kangaroo pump, it was noted that the programmed continuous feeds were set at a rate of 55cc/hr.The free water fluids programmed into the pump were set for 150cc every 4 hours.The hourly pump history did correctly reflect the continuous infusion but the flush history revealed that no free water flushes were administered automatically by the pump.
|
| |
|
Search Alerts/Recalls
|
|
|