BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V272
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
This cardiac resynchronization therapy pacemaker (crt-p) and left ventricular (lv) lead triggered a lead safety switch (lss) to unipolar due to a high out-of-range pace impedance measurement of 2,012 ohms.The vector was reprogrammed successfully and the lss for the lv lead was programmed off.This crt-p and lv lead remain in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
This cardiac resynchronization therapy pacemaker (crt-p) and left ventricular (lv) lead triggered a lead safety switch (lss) to unipolar due to a high out-of-range pace impedance measurement of 2,012 ohms.The vector was reprogrammed successfully and the lss for the lv lead was programmed off.This crt-p and lv lead remain in service.No adverse patient effects were reported.Additional information received reported that the cardiac resynchronization therapy pacemaker (crt-p) was explanted and replaced due to normal battery depletion (nbd).In addition, a left ventricular (lv) adapter was placed.This left ventricular (lv) lead remains in service.It was noted that the lv pace impedance measurement with the new crt-p was 1142 ohms.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTUA
Common Device Name
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15486097
Report Number2124215-2022-37411
Device Sequence Number8133072
Product Code NKE
UDI-Device Identifier00802526536663
UDI-Public00802526536663
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030005/S092
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/10/2016
Device Model NumberV272
Device Catalogue NumberV272
Device Lot Number100199
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer02/23/2023
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age54 YR
Patient SexFemale
-
-