| Model Number |
24705 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a rupture occurred.A 2.0mm x 220mm x 150cm coyote and a 4.0mm x 150mm x 150cm coyote were selected for use.During the procedure, the devices ruptured.There were no patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported that a rupture occurred.A 2.0mm x 220mm x 150cm coyote and a 4.0mm x 150mm x 150cm coyote were selected for use.During the procedure, the devices ruptured.There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a longitudinal tear 11.2cm from the tip and is approximately 48mm long.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|