| Model Number |
SRL-260 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/08/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: fold creases and the device was broken shell.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identified broken smooth.Based on the device analysis the final assessment is: a smooth edge broken in the side radius assessed as smooth opening.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Device has been explanted and replaced.
|
| |
|