MAUDE Adverse Event Report: ARJOHUNTLEIGH MAGOG INC. MAXI SKY 440; LIFT, PATIENT, NON-AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJOHUNTLEIGH MAGOG INC. MAXI SKY 440; LIFT, PATIENT, NON-AC-POWERED Back to Search Results
Model Number LE00000-AU
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Arjo was informed about the incident with the maxi sky 440 ceiling lift and easytrack fs installation system involvement.It was reported that the customer¿s staff transferred a patient from a bed to a powered wheelchair.After completing the transfer, the staff left the patient in the room.When the patient wanted to leave the room in the wheelchair, the easytrack fs frame partially fell forward twisting to the left.The patient managed to drive under the frame and left the room.No injury was reported as a consequence of the reported event.After the incident, the installation was inspected by an arjo service technician.The service technician observed that the lift attached to the track moved from one end to another due to frame collapse and that there were white marks on the installation track which came away when wiped.The technician placed the installation in the correct position and performed a function test.No problem occurred after the repair and the system was put back into service.
 
Additional Manufacturer Narrative
Arjo was informed about the incident with the maxi sky 440 ceiling lift and easytrack fs installation system involvement.It was reported that the customer¿s staff transferred a patient from a bed to a powered wheelchair.After completing the transfer, the staff left the patient in the room.When the patient wanted to leave the room in the wheelchair, the easytrack fs frame partially fell forward twisting to the left.The patient managed to drive under the frame and left the room.No injury was reported as a consequence of the reported event.After the incident, the installation was inspected by an arjo service technician.The service technician observed that the lift attached to the track moved from one end to another due to frame collapse and that there were white marks on the installation track which came away when wiped.The technician placed the installation in the correct position and performed a function test.No problem occurred after the repair and the system was put back into service.During the client's visit, the technician spoke with the patient involved in the incident and took photos showing the reported problem.According to the patient, the wheelchair was in such a position after completing the transfer that when she wanted to drive it out of the room, she drove it onto the right base of the installation.As a result, the frame twisted to the left, causing the entire installation to tilt forward.Based on the description of the event that occurred, we are unable to determine whether the interaction between the wheelchair on which the patient was located and the base of the installation could have resulted in the structure falling over.The instructions for use for the easy track fs (document number: 001.17300.En) instructs the user on how to make sure the installation is installed correctly through the following cautions: ¿caution: make sure both posts are correctly inserted into each base.¿ ¿caution: make sure the main hooks are well engaged.¿ ¿caution: never extend the post to allow the track to be in contact with the ceiling (.).¿ the document also indicates to: "lock the clamping collars on both posts to prevent the posts from extending by themselves.When the collars are locked, no red mark should be visible" and to: "adjust the levellers on each base to stabilize the bases.When adjusting the levellers, make sure the central gliders are in contact with the floor." the "care and maintenance" section of instructions for use for easytrack fs also contains the following information: "points to be inspected by user: before every use: make sure that bases are properly adjusted".Based on the information obtained from the service technician, the installation was installed on july 11, 2022, and the last inspection took place on august 10, 2022.No abnormalities in the operation of the device were found at that time.Based on the information obtained during the course of this investigation, we are unable to determine the exact root cause of the reported malfunction.To sum up, the maxi sky 440 ceiling lift and the easytrack fs installation were used as a system for the patient transfer when the installation frame collapsed and from that perspective, the system did not meet its performance specification.The complaint was decided to be reported to competent authorities due to the indication that the easytrack fs frame fell.No injury was sustained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXI SKY 440
Common Device Name
LIFT, PATIENT, NON-AC-POWERED
Manufacturer (Section D)
ARJOHUNTLEIGH MAGOG INC.
2001 tanguay street
magog, quebec J1X 5 Y5
CA  J1X 5Y5
Manufacturer (Section G)
ARJOHUNTLEIGH MAGOG INC.
2001 tanguay street
magog, quebec J1X 5 Y5
CA   J1X 5Y5
Manufacturer Contact
justyna kielbowska
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
883337089
MDR Report Key15486582
Report Number9681684-2022-00064
Device Sequence Number5256153
Product Code FSA
UDI-Device Identifier05055982795319
UDI-Public(01)05055982795319(11)220119
Combination Product (Y/N)N
Initial Reporter CountryNZ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberLE00000-AU
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Initial Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-