| Catalog Number |
2H8401 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a clearlink system non-dehp solution set broke below the lower y-site while priming with normal saline.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph using the naked eye which observed the tubing was separated from the y-site clealink in the upstream part.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause is due to improper manual assembly during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|