DEPUY SPINE INC CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 13G X 4 INCHES; NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE
|
Back to Search Results |
|
| Model Number |
283903413 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Part/lot combination is not valid, therefore the dhr could not be completed.If the device is returned or the lot number is confirmed the dhr will be revisited.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in colombia as follows: it was reported that on (b)(6) 2022, the needle would not come out after the application of cement.The surgeon tried to remove it by himself, and the needle broke.The broken fragment was taken out through the laminectomy incision.An x-ray was performed, and no fragments were seen in the vertebra.The patient had a metastatic tumor on the spine with hard bone.The surgeon passed the needle until the middle of the vertebral body and could not advance it any further, so he decided to inject confidence spinal cement.Surgery was delayed because they had to pass other equipment to be able to remove the fragment.This report involves one confidence spinal cement system introducer needle, diamond tip 13g x 4 inches.This is report 1 of 1 for (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|