Model Number 787426
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were burrs at the tip of the inlay optima ureteral stent, making it unusable.The stent was not used and they inserted a new one.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that there were burrs at the tip of the inlay optima ureteral stent, making it unusable.The stent was not used and they inserted a new one.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.The ureteral stent was returned without the original packaging.An evaluation was completed.Photo sample was provided and the tip of the stent appeared to be mangled.The tips of the physical sample were evaluated and could be seen as defective tips.The first tip evaluated appears the tubing is twisted and caused the tube to curl up and stretch the material.The second tip evaluated appears to be going inward, there was also indentations indicating the sample was being clamped with a foreign object.Based on the appearance of the sample it could be confirmed that the sample would not pass 100% inspection in this state.The defect at the tips could not be recreated, and no defect in this manner has been seen at this facility.The ends where the defect occurred did appear to have black smudging on the outside of the stent.The dimensional requirements could not be 100% executed due to the state of the sample.The outer diameter measured 0.0621" using a laser micrometer which was within specifications.The pin gauge could not be inserted to measure the inner diameter.The guidewire for passage could not be inserted due to the damage at the tips of the sample.Although a specific cause cannot be determined, a potential root cause for this event could be, "user does not handle with care, bends sharply".As no patient involvement was reported the device was not used, however, it is intended for treatment purposes.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15487707
Report Number1018233-2022-07459
Device Sequence Number13942542
Product Code FAD
UDI-Device Identifier10801741015363
UDI-Public(01)10801741015363
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2026
Device Model Number787426
Device Catalogue Number787426
Device Lot NumberNGGQ5107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer02/16/2023
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date02/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-