Model Number CB6004
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Fill volume: 750 ml, flow rate: 10ml/hr , procedure: unknown, cathplace: unknown.It was reported the 750ml on-q pump emptied out in less than 42-hours with the rate set at 10ml/hr.There was no reported injury.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 26-sep-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 80401100, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The device was evaluated.The pump was received empty, it was filled to nominal volume with saline, flow accuracy test was performed, and the pump yielded a rate within specifications, pressure pot testing was performed on the documented saf (select-a-flow) settings, and all tested rates met specifications, the fast flow reported by the customer was not replicated during the functional evaluation.The investigation determined that the returned device functioned as intended during the sample evaluation, no defect was found; therefore, complaint is not confirmed.Root cause could not be determined.All information reasonably known as of 16-nov-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q PAIN RELIEF SYSTEM WITH SELECT-A-FLOW, 600 ML, 2-14 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15488129
Report Number2026095-2022-00107
Device Sequence Number1856640
Product Code MEB
UDI-Device Identifier00193494134761
UDI-Public00193494134761
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/20/2023
Device Model NumberCB6004
Device Catalogue NumberN/A
Device Lot Number80401100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date09/26/2022
Supplement Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROPIVACAINE 0.2%; UNKNOWN CATHETER
Patient Age75 YR
Patient SexMale
Patient Weight94 KG
-
-