• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AROA BIOSURGERY LTD. OVITEX REINFORCED TISSUE MATRIX; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AROA BIOSURGERY LTD. OVITEX REINFORCED TISSUE MATRIX; SURGICAL MESH Back to Search Results
Model Number F10244-1020P
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the relevant batch record was conducted which showed the device lot met all release specifications and no relevant anomalies or non-conformances were noted.No other similar reports have been noted associated with this lot aside from these four recurrences noted.A review of sales records noted that this device lot has been utilized at this hospital 34 times with no other issues reported.Further, ball burst testing was completed on available samples from the same lot.All samples tested met the pre-defined strength specification.It was noted by the surgeon that the hernia recurrence repair was conducted in a different manner as the original repair, thus a surgical assessment of the original repair and the surgical mesh could not be conducted.
 
Event or Problem Description
A surgeon reported four separate indirect inguinal hernia recurrences after inguinal hernia repair with ovitex core.All recurrences were addressed via a second repair operation.This is one of these four reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVITEX REINFORCED TISSUE MATRIX
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
AROA BIOSURGERY LTD.
2 kingsford smith place
airport oaks, auckland 2022
NZ  2022
Manufacturer (Section G)
AROA BIOSURGERY LTD.
2 kingsford smith place
airport oaks, auckland 2022
NZ   2022
Manufacturer Contact
tina o'brien
2 kingsford smith place
airport oaks, auckland 2022
NZ   2022
MDR Report Key15489352
Report Number3007321028-2022-00038
Device Sequence Number1919038
Product Code FTM
UDI-Device Identifier09421904065031
UDI-Public09421904065031
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K153632
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberF10244-1020P
Device Catalogue NumberF10244-1020P
Device Lot NumberERT-21K10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Initial Report FDA Received Date09/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-