| Model Number |
2490-36-000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the femoral impactor chipped off the black impactor.Broken piece was removed from the field and disposed.No surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this complaint was returned for examination.Visual examination found brakeage of the femoral impactor head, confirming the reported allegation.The broken part was not returned.Additionally, the overall condition of the device heavily worn with usage marks all over the device.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Drawing/specifications reviewed: (b)(4).Document reviewed: wi (b)(4)| rev:3.Device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the investigation was not able to retrieve the required lot code.
|
| |
|
Search Alerts/Recalls
|