• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0480-01C
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complete event site name: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id#: (b)(4).
 
Event or Problem Description
It was reported that after approximately 30 hours of intra-aortic balloon (iab) therapy the console generated a blood detected alarm and blood was seen in the tubing.The iab was removed and a blood clot was found in the iab.A new iab was not inserted; treatment was continued with medication.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and traces of blood on the exterior of the catheter.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and a leak was detected on the membrane approximately 1.5cm from the rear seal measuring 0.013cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key15490851
Report Number2248146-2022-00746
Device Sequence Number1995773
Product Code DSP
UDI-Device Identifier10607567109343
UDI-Public10607567109343
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2024
Device Model Number0684-00-0480-01C
Device Catalogue Number0684-00-0475
Device Lot Number3000208097
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN PUMP
Patient Age82 YR
Patient SexMale
Patient Weight65 KG
-
-