Model Number 0998-00-3013-85
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during routine check, the cs100 intra-aortic balloon pump (iabp) helium was leaking from the unit.There was no patient involvement.
 
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Additional Manufacturer Narrative
Updated fields: b4, d9, e1(site country, email address), e2, e3, g3, g6, h2, h3, h4, h6 ( type of investigation, investigation findings, investigation conclusion, component code), h10, h11.Corrected fields: g2.A getinge field service engineer evaluated helium leak from unit after checking leakage found from the tubing which was connected with the helium regulator assembly same needs for testing purpose further checking will be done after replacing the same.The helium multiple was replaced.Unit now working fine and handover to customer with working condition.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 UTS INTRA-AORTIC BALLOON PUMP INTL 220V
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15490855
Report Number2249723-2022-02472
Device Sequence Number1996264
Product Code DSP
UDI-Device Identifier10607567108094
UDI-Public10607567108094
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K112372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3013-85
Device Catalogue Number0998-00-3013-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-