• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) INSYTE-W WINGED PNK 20GA X 1.88IN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) INSYTE-W WINGED PNK 20GA X 1.88IN; INTRAVASCULAR CATHETER Back to Search Results
Model Number 381337
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using the insyte-w winged pnk 20ga x 1.88in, the catheter broke.The following information was provided by the initial reporter: catheter initially appearing to be kinked, line not sutured.Upon repositioning catheter and removing the securing strip over the hub, catheter noted to be completely broken off at the hub, with the catheter still in the vessel (pusatile blood flow noted).
 
Additional Manufacturer Narrative
Medical device expiration date: unknown.Initial reporter facility name: (b)(6) health sciences centre / (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
H6: investigation summary: one photo was received by our quality team for evaluation.From the photo, the catheter tubing was observed to be broken off near the nose of the adapter.A slanted cut was observed at the breakage point.A device history record could not be evaluated as the lot number is unknown.The assembly process was reviewed, the only station that is in contact with the catheter tubing is the rubber pads at the flaring station.Given the elastic nature of the rubber pads and that there are no sharp or hard edges on the pads, this station is unlikely to cause the observed slanted cut on the catheter tubing.The slanted cut on the catheter could occur during product application when the product was manipulated.Current quality controls include a daily outgoing functional test and a start of shift in-process functional test in place to check for catheter-to-adapter pull force.There is also a daily outgoing visual inspection and in-process visual inspection in place to check for catheter damage.The actual root cause cannot be established.
 
Event or Problem Description
It was reported that while using the insyte-w winged pnk 20ga x 1.88in, the catheter broke.The following information was provided by the initial reporter: catheter initially appearing to be kinked, line not sutured.Upon repositioning catheter and removing the securing strip over the hub, catheter noted to be completely broken off at the hub, with the catheter still in the vessel (pusatile blood flow noted).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSYTE-W WINGED PNK 20GA X 1.88IN
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15491488
Report Number8041187-2022-00562
Device Sequence Number16608362
Product Code FOZ
UDI-Device Identifier30382903813378
UDI-Public30382903813378
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number381337
Device Catalogue Number381337
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-