| Model Number |
7209686 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, in the joint cavity, the skin of the probe saphyre wand crack at the beginning of the ablation.The procedure was completed with a s+n back up device.There was a non-significant surgical delay and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection found the device was returned without original packaging.The insulation at the distal tip was split and pushed out from one point.The return electrode had debris.A functional evaluation found the insulation did not crack or peel further when operated in cut or coagulation mode.The complaint was confirmed and the root cause associated with an unintended use of the device.Factors that may have contributed to the reported event include using the probe as a lever in the joint space or contact with another instrument during the procedure.No containment or corrective actions are recommended at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).Corrected data: h1 to malfunction.
|
| |
|