Model Number 5620
Medical Device Problem Codes Premature Discharge of Battery (1057); Failure to Interrogate (1332)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an in clinic follow up, the device was unable to be interrogated.It was suspected the device had reached end of service.The patient was stable.Further information has been request but has not yet been provided.
 
Event or Problem Description
Additional information was received indicating the device was explanted and replaced to resolve the event.The patient was stable.
 
Additional Manufacturer Narrative
The field complaint of premature discharge of battery was not confirmed.The reported event of failure to interrogate was confirmed.The device was received unable to be interrogated.The device was cut open and it was determined that battery voltage was at end of life.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Functional testing was performed with new battery installed, and the device was found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR SR
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15492096
Report Number2017865-2022-39415
Device Sequence Number1958029
Product Code NVZ
UDI-Device Identifier05414734501606
UDI-Public05414734501606
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2014
Device Model Number5620
Device Catalogue Number5620
Device Lot Number3974449
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer09/28/2022
02/20/2023
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date10/12/2022
03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-