ABBOTT MEDICAL EON C 16-CHANNEL IPG; SCS IPG Back to Search Results
Model Number 3688
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient's ipg turned inoperable and the patient lost therapy.It is unknown of the ipg depleted prematurely.Surgical intervention may occur to address the issue.
 
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Additional information received indicated the patient underwent surgical intervention wherein the ipg was explanted and replaced to address the issue.During the procedure, it was noted the lead was disconnected from the patient's ipg.The new ipg was connected to the existing lead and therapy was restored post-operatively.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain patient weight.A patient experiencing an inoperable ipg was reported to abbott.The patient¿s ipg was explanted and replaced.Therapy was restored.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON C 16-CHANNEL IPG
Common Device Name
SCS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15492511
Report Number1627487-2022-05171
Device Sequence Number11226600
Product Code LGW
UDI-Device Identifier05414734402460
UDI-Public05414734402460
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2013
Device Model Number3688
Device Catalogue Number3688
Device Lot Number3416678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR X2; SCS LEAD
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-