DEPUY SYNTHES PRODUCTS LLC 3MM FBALL,ST,M; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Model Number |
M-3SB-G1 |
| Medical Device Problem Codes |
Dull, Blunt (2407); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by switzerland that during service and evaluation, it was determined that the burr device cutting edges were broken.It was further determined that the device failed visual inspection and some traces of corrosion were identified on the shaft during visual assessment.It was noted in the service order that the device did not function while in use with an attachment device.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: concomitant med products and therapy dates: attachment device, (b)(6) 2022.Device evaluation: the actual device was returned for evaluation.Quality engineering evaluated the device and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from wear.Device history record review: a device history review was performed and no non-conformances were identified related to the reported condition.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|