• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 21.5 CM; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 21.5 CM; HIP PROSTHESIS Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision surgery approximately two (2) months post-op due to implant fracture.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Exact implant date is unknown.The device was implanted in (b)(6) 2022.Report source - canada.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The nail was returned for investigation.It is possible to confirm the event, as the nail is fractured at the level of the hole where the lag screw is inserted.The outer surface of the nail is inconspicuous with very little signs of damage.Inside the hole for the lag screw, it is possible to see signs of wearing, probably coming from the lag screw contacting the surface.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Op notes of the surgery were not provided therefore it is not possible to assess the surgical technique.Complaint description reports a possible fall injury of the patient that led to the fracture of implant; however, the event is not confirmed.With the available information a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 21.5 CM
Common Device Name
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15492985
Report Number0009613350-2022-00496
Device Sequence Number1780047
Product Code HSB
UDI-Device Identifier00889024298873
UDI-Public(01)00889024298873(17)320117(10)3096325
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2493-211-10
Device Lot Number3096325
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-