• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 4 6MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 4 6MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 1517-10-406
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient required first stage right total knee revision for a fungal infection of the right knee.Pt already had in situ attune revision implants, requiring full extraction of the prosthesis.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS RP INSRT SZ 4 6MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15493792
Report Number1818910-2022-18785
Device Sequence Number11234107
Product Code NJL
UDI-Device Identifier10603295055792
UDI-Public10603295055792
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number1517-10-406
Device Catalogue Number151710406
Device Lot Number9773459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/14/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL RT SZ 4 CEM; ATTUNE CRS FEMORAL RT SZ 4 CEM; ATTUNE DIST FEM AUG SZ 4 12MM; ATTUNE DIST FEM AUG SZ 4 12MM; ATTUNE DIST FEM AUG SZ 4 12MM; ATTUNE DIST FEM AUG SZ 4 12MM; ATTUNE POS FEM AUG SZ 4 8MM; ATTUNE POS FEM AUG SZ 4 8MM; ATTUNE POS FEM AUG SZ 4 8MM; ATTUNE POS FEM AUG SZ 4 8MM; ATUN FEM SLV M/L 40MM FULL POR; ATUN FEM SLV M/L 40MM FULL POR; ATUN PRESSFIT STR STEM12X110MM; ATUN PRESSFIT STR STEM12X110MM; ATUN PRESSFIT STR STEM12X110MM; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM12X60MM; ATUNE PRESSFIT STR STEM12X60MM; ATUNE REV RP TIB BASE SZ 3 CEM; ATUNE REV RP TIB BASE SZ 3 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-