• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AURIS HEALTH, INC. MONARCH PLATFORM; BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AURIS HEALTH, INC. MONARCH PLATFORM; BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES Back to Search Results
Model Number MON-000005-01
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer (fse) went onsite performed a physical inspection of the camera pathway, performed a functional check of the vision pathway using a scope and was not able to replicate the issue.As such, a cause could not be determined.No review of the device history record (dhr) is necessary as the issue could not be confirmed or replicated.A review for similar reported issue notes this to be a known issue but has not established a possible root cause.The risk associated with the reported issue is noted in: risk document number: 105-000367-00, rev.M, dfmea bronch 2.0 risk id: fmea-1240 & fmea-0510.
 
Event or Problem Description
It was reported the physician lost the live camera image after completing the navigation registration and started to drive out to the target.The physician proactively power cycled the system and changed out the scope; however the issue persisted.The physician was unable to perform the monarch procedure due to the system related issue, the case was aborted and rescheduled for a later date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH PLATFORM
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES
Manufacturer (Section D)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer (Section G)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer Contact
harminder kaur
150 shoreline drive
redwood city, CA 94065
5106057113
MDR Report Key15494493
Report Number3014447948-2022-00034
Device Sequence Number12057548
Product Code EOQ
UDI-Device IdentifierB634MON000005010
UDI-Public+B634MON000005010/16D20210706
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K211493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMON-000005-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Initial Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-