| Catalog Number |
STP-PR-001 |
| Medical Device Problem Code |
Premature End-of-Life Indicator (1480)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that early sensor expiration occurred.The sensor was inserted into the abdomen no product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Cmpl-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that early sensor expiration occurred.The sensor was inserted into the abdomen.Data was evaluated and the allegation was confirmed.The probable cause was determined to be high counts aberration via data.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|