| Model Number |
1516-20-810 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for instability.Event is serious and is considered moderate.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2021.Date of revision: (b)(6) 2022.Date of event: (b)(6) 2022.(left knee).Treatment: revision; insert was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Added: b5, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient developed moderate lateral laxity with pain and swelling and underwent a revision of the insert while the knee was also rebalanced on (b)(6) 2022.Intraoperatively, scarring was found in the anterior compartment and 2+ to 3+ lateral laxity.Extensive soft tissue release was performed and a 4mm thicker insert was placed with a satisfactory range of motion.No complications were noted.Doi: (b)(6) 2021 (left knee).Dor: (b)(6) 2022 (left insert only).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|