| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
|
| Date of Event |
08/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was explanted from the left eye due to severe halos, glare, and lack of near vision.A non-johnson & johnson 3-piece iol was implanted as replacement.It is unknown if there was any patient injury or if medical or surgical intervention was required.The patient is stable post-operatively with the new lens, with a 3+ posterior capsule opacification (pco).The device was discarded.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event: unknown; requested but not provided.The best estimate date is between aug 2019 and aug 29, 2022.If implanted, give date: unknown, information was requested but not provided.The best estimate date is aug 2019.The device was not returned for evaluation and the serial number for this device is unknown/not provided; therefore, no further product investigation can be performed.Should any further relevant information become available a supplemental medwatch will be filed.Device manufacture date: unknown, as the serial number was not provided.Multiple attempts have been made to obtain additional information regarding the event; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: through follow-up, it was learned that the patient's original date of implant was (b)(6) 2019 and the symptoms were first noted on (b)(6) 2019.The patient's pre-operative corrected near vision visual acuity (va) was j1+, pre-operative uncorrected va was j3.The patient's post-operative corrected near vision visual acuity (va) was not checked, post-operative uncorrected va was j6.Th patient was last seen on (b)(6) 2020 and at the time the patient's spectacle corrected distance visual acuity (scdva) was 20/30 and the mrx showed -0.50+0.50x120 and va was 20/20.The spherical equivalent of the mrx would be -0.25sph and the surgeon aims for -0.15 to -0.25.No further information was provided.The following sections have been updated accordingly: section a5: ethnicity: not hispanic/latino section b3: date of event: (b)(6) 2019.Section d6a: if implanted, give date: (b)(6) 2019.Section d4: model number: zxr00.Section d4: serial number: (b)(6).Section d4: catalog number: zxr00u0205.Section d4: expiration date: aug 2, 2023 section d4: unique identifier (udi) number: (b)(4).Section h4: device manufacture date: aug 2, 2018.Section h3-other (81): the device was not returned for evaluation as it was discarded; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|