Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) was explanted from patient's right eye due to blurry vision.The pre-operative visual acuity was 20/30 and post-implantation with the original lens, it was 20/50 uncorrected and best corrected visual acuity was 20/29 (initially indicated as -1.0 20/20).Another lens of same model and diopter was used as the replacement.Post-explant, the visual acuity was 20/25 and the patient was doing well.The incision was enlarged to remove the iol from the eye but no vitrectomy required and no sutures were done.There was no delay in procedure reported.No other information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt was made to obtain missing information; however, to date, the information was not available.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information/corrected data: section b5: additional information received clarified that the best corrected visual acuity post-implantation with the original lens was -1.0 20/20 (as indicated initially) and not 20/29.In review, it was also noted that fields of "d2" of the initial mdr report were inadvertently populated with incorrect information therefore, the corrections have been included in this supplemental mdr report and the following fields were updated accordingly: section d2: common device name: lens, multifocal intraocular.Section d2: device product code: mfk.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: oct 13, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint lens was received cut, but not cut in half.The lens was cleaned and, no issues that could contribute to or cause blurry vision.Based on the return condition of the lens, no further inspection can be performed.Conclusion: the complaint issues were not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15495029
Report Number3012236936-2022-02399
Device Sequence Number5856402
Product Code MFK
UDI-Device Identifier05050474750739
UDI-Public(01)05050474750739(17)241019
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/21/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/19/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU230
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer10/12/2022
10/19/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date10/13/2022
10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;