| Model Number |
2545-05-510 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The surgeon informed via email that he felt his mto im drill guide positions the im hole too low for larger patients, so that on a size 9 and 10 balance block, there is not enough excursion of the threaded knob before it disengages.It is not clear whether this event is a direct observation from surgery, or something that was noticed outside of surgery on a foam bone.No patient harm is noted as unclear if patient involved.Note ¿ from the description, this would appear to be an observation based on the design rather than something that is batch or lot specific.There is no information on how many items are involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The investigation could not verify or draw any conclusions about the root cause of the reported event without the device to examine.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) was not possible, because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|