| Model Number |
777624 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the front tip of the ureteral stent was found to be broken.Therefore, a new stent was unpacked to use.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.Although the specific cause cannot be determined, based on the risk document a potential root cause could be, "inappropriate package design".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "improper handling technique can seriously weaken the stent." "exercise care.Tearing of the stent can be caused by sharp instruments." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." correction: g h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the front tip of the ureteral stent was found to be broken.Therefore, a new stent was unpacked to use.
|
| |
|
Search Alerts/Recalls
|