ANGIODYNAMICS BIOFLO DIALYSIS CATHETER; CATHETER, HEMODIALYSIS, IMPLANTED Back to Search Results
Catalog Number 10302804
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was noted there was a leak in the extension of the duramax bioflo catheter in the venous branch, at the junction point of the transparent extension with the luer.The catheter was removed and will be replaced.The patient did not experience any adverse effects, harm, or require medical intervention because of this incident.
 
Additional Manufacturer Narrative
Returned for evaluation was a bioflo dialysis catheter.As received, it was noted during the visual inspection that there was a hole just below the female luer {arterial red clamp side}.The reported complaint device failure mode of catheter leaked was confirmed.Hole was observed in the arterial extension tube adjacent to the flat section of the hub, i.E.90° from the parting line.There are no visible indications that the failure is related to either the manufacturing process or associated tooling.The most likely root cause for this failure is over torquing of the hub luer-to-extension tubing junction during cleaning/use of the device.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the following is provided as a reference from dfu: warnings in the rare event that a hub or connector separates from any component during insertion or use, take all necessary steps and precautions to prevent blood loss or air embolism and remove catheter.Use of excessive pull force on the catheter may cause the suture wing to detach from the bifurcate.Do not use acetone on any part of the catheter tubing.Exposure to this agent may cause catheter damage.Do not use sharp instruments near the extension tubing or catheter lumen.Do not use scissors to remove dressing.Catheter will be damaged if clamps other than what is provided with this kit are used.Clamping of the tubing repeatedly in the same location may weaken tubing.Avoid clamping near the luers and hub of the catheter.Examine catheter lumen and extensions before and after each treatment for damage.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO DIALYSIS CATHETER
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15496123
Report Number1317056-2022-00118
Device Sequence Number12272304
Product Code MSD
UDI-Device IdentifierH965103028041
UDI-PublicH965103028041
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K131260
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number10302804
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Initial Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-