It was reported that, during a navio assisted tka surgery, it was found that the navio handpiece thumberscrew is fractured.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed.
|
|
Additional information: d8, h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: the navio handpiece thumberscrew, part number pfsr100026, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with repeated over torquing resulting in thumbscrew shaft fracture.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|