|
AVANOS MEDICAL INC. MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE, ORAL/NASAL MAGILL-5.0-3.5-12; VAP MICROCUFF ENDOTRACHEAL TUBES
|
Back to Search Results |
|
| Model Number |
35112 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Avanos medical received a single report that referenced three different incidences, which were associated with separate units, involving three different events.This is the third of three reports.Refer to 3011270181-2022-00072 for the first event.Refer to 3011270181-2022-00073 for the second event.It was reported that the endotracheal tube (ett) cuff deflated during use.The ett tube was replaced with a new tube.No further medical interventions reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 10-nov-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Additional information received 10-nov-2022 stated "as the batch has been confirmed as the same batch as the end-user reported.In addition, the end-user reported on the 01-sep-2022 "the consultant changed the ett tube again today due to the cuff leaking again overnight.".
|
| |
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.The sample provided was the actual sample for the second incident with the same lot number (cm9336003).The device history record for the reported lot number, cm9336003, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 12-jan-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|
|
|