BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00516700
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code Chest Pain (1776)
Date of Event 09/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation on september 06, 2022 that a wallflex esophageal stent was implanted in the esophagus to treat a 1.5 cm tracheoesophageal fistula during esophageal stent placement procedure performed on (b)(6), 2022.Reportedly, the patient's anatomy was not tortuous and was dilated from the teeth to esophagus with 20-22 cm dilation prior to stent placement.Post stent placement, the patient experienced chest pain.Pain medication was administered but the pain did not subside so the patient was scheduled for a stent removal procedure.On (b)(6), 2022, during stent removal procedure, it was noted under endoscopic view that the stent retrieval loop was broken.The physician attempted to remove the broken stent but was unsuccessful because the stent had no retrieval loop.Reportedly, the patient passed away due to esophageal cancer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15499308
Report Number3005099803-2022-05610
Device Sequence Number14719885
Product Code ESW
UDI-Device Identifier08714729778035
UDI-Public08714729778035
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K091510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00516700
Device Catalogue Number1670
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/06/2022
Initial Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-