Model Number B1080-040
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified, mildly tortuous, 90% lesion in the right external iliac artery.Atherectomy was performed and an 8x40mm armada 35 balloon dilatation catheter (bdc) was advanced to the target lesion and inflated one time; however, it ruptured at burst pressure.Subsequently, an 8x60mm armada 35 bdc was advanced and inflated one time to nominal pressure when the balloon ruptured and separated from the catheter in the right external iliac artery.One hour was spent trying to snare out the separated segment, however it was unsuccessful.An 8.X5cm non-abbott covered stent was implanted to embed the separated segment in the artery.There was no adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.The lot history record (lhr) for this product could not be reviewed and a similar complaint query could not be performed because the product was not returned for evaluation and the lot number was not reported.Based on the information provided, the reported material rupture appears to be due to circumstances of the procedure.It is likely that the balloon rupture occurred due to interaction with calcification.There was no report of leaks noted during preparation for use, suggesting that the balloon damage was not pre-existing.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The 8x60mm armada 35 balloon referenced is filed under separate medwatch report number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15499392
Report Number2024168-2022-10079
Device Sequence Number17438599
Product Code LIT
UDI-Device Identifier08717648154867
UDI-Public08717648154867
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB1080-040
Device Catalogue NumberB1080-040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2022
Initial Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VIABHAN STENT
Patient Age65 YR
Patient SexMale
-
-