C.R. BARD, INC. (BASD) -3006260740 STATLOCK IV SELECT MACRO 6IN ULTRASITE 25 BOX; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of jugp0232 showed no other similar product complaint(s) from this lot number.Device not returned.
 
Event or Problem Description
It was reported " fluid leaked from a crack in the hub.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a crack in the luer adaptor was inconclusive due to the sample condition.A photograph of a 6¿ macro extension tubing set from an iv select statlock device was returned for evaluation.An arrow was drawn near the sample which pointed to the female luer adaptor, which is where the sample reportedly leaked from.No obvious cracks or deformation was noted in the luer adaptor.However, no close-up photographs of the reported damage were provided and therefore the damage may not have be visible to the investigator.There were no distinguishing features in the provided photograph which could be used to independently confirm the complaint and/or aid in identification of a specific root cause.Possible contributing factors for a cracked luer adaptor could include dimensional incompatibility, forceful insertion of the valved connector during manufacturing, damage from shipping/handling, or other unidentified factors.A review of the manufacture quality and inspection records for the specific lot involved found no potentially related issues encountered during the manufacture and assembly of that product lot.A lot history review (lhr) of jugq3281 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported " fluid leaked from a crack in the hub.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK IV SELECT MACRO 6IN ULTRASITE 25 BOX
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15499396
Report Number3006260740-2022-03776
Device Sequence Number5666534
Product Code FPA
UDI-Device Identifier00801741082115
UDI-Public(01)00801741082115
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K980992
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2025
Device Model NumberN/A
Device Catalogue NumberIV0572
Device Lot NumberJUGQ3281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date09/28/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-