| Model Number |
SXPD2B402 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records for the finished good lot and the raw material components identified no quality issues during the incoming, manufacturing, in-process or final inspection processes.The needle component is supplied by ethicon.No samples or magnified photos were returned for testing or review.No retained samples are available for testing/review.The third-party report was not received to date.If samples or photos become available at a later time the devices will be evaluated, photos reviewed and the results will be included in a follow-up report.A potential root cause for a needle detaching when slightly tugged when preparing for use or detaching during the procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible that the needle device is being gripped on or near the swaged area, causing damage to the attachment, which allows the needle to pull free from the suture.Needles have detached from the suture when excessive force was applied, exceeding the strength of the attachments.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point¿.Without receiving the actual device for review, receiving magnified photos of the device, receiving devices from the same finished good lot to test/review, receiving the results of the third party evaluation or receiving detailed information regarding the preoperative preparation of the device, tools utilized to grasp the device, the procedure performed or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
|
| |
|
Event or Problem Description
|
|
Our distributor is reporting the during the suturing process, the suture''s needle and line separated.The surgery was delayed.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
No sterile samples or magnified photos were returned for testing or review.No retained samples were available for review.If samples or photos become available at a later time the devices will be evaluated, photos reviewed, and the results will be included in the file.A revised response will be forwarded to you at that time.The actual device was returned with the foil package labeled as lot aady524.The needles and suture appeared used (ex.Blood, tool marks, engaged barbs).Tool marks were observed on the swage end and near the tip of the needles; one (1) tip was bent.The needles remained securely attached to the suture pieces when gently tugged during the analysis.The end point of failure on the 1st piece of suture appeared smooth, slightly flared, cut with a sharp surgical instrument.The end point on the 2nd piece of suture appeared elongated, sheared up the side of the suture.The barbs near the end point are damaged.The barbs along the suture strand are engaged; no unusual damage observed.It appears a segment of the suture device was not included in the package.A potential root cause for a needle detaching when slightly tugged during the procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible that the device is being grasped with surgical instruments, damaging the device, allowing the suture to break during use.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point.¿ without receiving the entire device for review, receiving magnified photos of the entire device for review, receiving devices from the same finished good lot to test/review, or receiving detailed information regarding the preoperative preparation of the device, tools utilized to grasp the device, procedure performed, method utilized to secure the device in the tissue or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
|
| |
|
Search Alerts/Recalls
|
|