BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG80166
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly dislodged.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, an image was provided for review.The investigation of the reported event is currently underway.Expiry date: 06/2025.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the balloon markers were allegedly dislodged.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one atlas gold pta dilatation catheter was received for evaluation.During visual evaluation, an x-ray picture was taken of the balloon showing the marker bands ranging from a distance of 30mm.During functional evaluation, the balloon was cut in order to see the condition of the marker bands.The measurements were 6.0 and 9.0cm from the distal tip.No anomalies were noted to both bands.No other functional testing was performed.One fluoroscopic image was provided and reviewed.The image centered around the left pelvis was assessed.A stent with an inflated angioplasty balloon was visualized.It appeared that the balloon was dislodged from its delivery catheter.Therefore, the investigation is confirmed for the reported marker bands dislodgement as the marker bands were outside of the specification mentioned in the design and development document.Additionally, the provided fluoroscopic image confirms the dislodgement.A definitive root cause for the reported marker bands dislodgement could not be determined based upon the provided information.H10: d4 (expiry date: 06/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15500306
Report Number2020394-2022-00768
Device Sequence Number19331632
Product Code DQY
UDI-Device Identifier00801741060830
UDI-Public(01)00801741060830
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG80166
Device Catalogue NumberATG80166
Device Lot Number93TG0130
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date09/28/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-