• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL TECHNOLOGY, INC. MINI ONE(R) BALLOON BUTTON; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL TECHNOLOGY, INC. MINI ONE(R) BALLOON BUTTON; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number M1-5-1417-I
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is a response to uf report # (b)(4) which was received by amt on 09/06/2022.Based on the provided information, the complaint is not a reportable event per 21 cfr section 803.There was no death.There was no serious injury, as defined by the fda.This was not life-threatening, nor did this result in permanent impairment or necessitate medical or surgical intervention to preclude permanent impairment.The device wasn't opened or used, as it was returned to amt for examination still sealed.Amt confirmed through the examination of the returned device that a hair was partially wrapped around the button of the device.A device history review was completed for the reported lot number and no anomalies were found and there have been no other complaint reports from this same batch.It is believed that the hair was hidden under a component in the tray during the multiple qc inspections of the product prior to shipment, and moved into a visible area of the tray during shipment.Production and warehouse staff have been notified accordingly of the hair to ensure repeat occurrence is avoided.The complaint information has been logged into our complaint database for trending purposes.Complaint # (b)(4) was assigned to this report.
 
Event or Problem Description
Per the original reporter in uf/importer report #: (b)(4), " a sterile package with a 14fr x 1.7cm mini one(r) balloon button (boxed kit - isosaf) low profile feeding device that was unopened was found to have a hair that was wrapped around the button portion in the kit.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI ONE(R) BALLOON BUTTON
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine boulevard
brecksville OH 44141
Manufacturer (Section G)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine boulevard
brecksville OH 44141
Manufacturer Contact
joshua meinke
8006 katherine boulevard
brecksville, OH 44141
MDR Report Key15501344
Report Number1526012-2022-00011
Device Sequence Number5439081
Product Code KNT
UDI-Device Identifier00842071113971
UDI-Public(01)00842071113971(17)241201(10)220104-426
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K161413
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2024
Device Model NumberM1-5-1417-I
Device Catalogue NumberM1-5-1417-I
Device Lot Number220104-426
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2022
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/06/2022
Initial Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-