CROSSROADS EXTREMITY SYSTEMS CROSSTIE INTRAOSSEOUS FIXATION SYSTEM; SCREW, FIXATION, BONE Back to Search Results
Model Number 1443-2511
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Crossroads received a broke crosstie implant that was still lodged in the inserter on (b)(6) 2016.Contact was made with event reporter and it was discovered that the implant broke during insertion.Another crosstie kit was opened and successfully used.Broken fragments of implant were removed.The lhr was reviewed for p/n 1443-2511 lot 101471.The implant lhr was also reviewed (p/n 100060 lot 101442).No evidence was found to suggest the device or the implant did not meet specification prior to distribution.
 
Event or Problem Description
Crossroads received a broke crosstie implant that was still lodged in the inserter on (b)(6) 2016.Contact was made with event reporter and it was discovered that the implant broke during insertion.Another kit was opened and successfully used.Broken fragments of implant were removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSTIE INTRAOSSEOUS FIXATION SYSTEM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
CROSSROADS EXTREMITY SYSTEMS
6423 shelby view dr.
suite 101
memphis TN 38134
Manufacturer Contact
6423 shelby view dr
suite 101
memphis, TN 38134
9012218406
MDR Report Key15501423
Report Number3020584246-2022-00007
Device Sequence Number15392345
Product Code HWC
UDI-Device Identifier00815432022052
UDI-Public00815432022052
Combination Product (Y/N)N
PMA/510(K) Number
K151270
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1443-2511
Device Catalogue Number1443-2511
Device Lot Number101471
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2016
Initial Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age14 YR