| Model Number |
1011-0139-04 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the associated device failed to discharge using these internal handles.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The internal handles will not be returned to zoll medical corporation.The biomed reported to zoll medical corporation that an internal inspection found a cut in the cable insulation showing internal wiring.Due to the age of these internal handles (47 months), they have most likely met their useful life and exceeded the maximum of 50 autoclave/sterilization cycles.The customer will remove this set of internal handles from service and will order a replacement set.Analysis of reports of this type has not identified an increase in trend.
|
| |
|