• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CROSSROADS EXTREMITY SYSTEMS CROSSTIE INTRAOSSEOUS FIXATION SYSTEM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CROSSROADS EXTREMITY SYSTEMS CROSSTIE INTRAOSSEOUS FIXATION SYSTEM; SCREW, FIXATION, BONE Back to Search Results
Model Number 1443-2501
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Both the kit (p/n 1443-2501 lot 101048) and implant (p/n 100059 lot 100932) lhrs were reviewed and no evidence was found to suggest that the device did not meet specifications prior to distribution.
 
Event or Problem Description
Implant broke while doctor was attempting to insert the implant proximally.A k-wire was used and implant removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSTIE INTRAOSSEOUS FIXATION SYSTEM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
CROSSROADS EXTREMITY SYSTEMS
6423 shelby view dr.
suite 101
memphis TN 38134
Manufacturer Contact
6423 shelby view dr
suite 101
memphis, TN 38134
9012218406
MDR Report Key15501626
Report Number3020584246-2022-00015
Device Sequence Number12083719
Product Code HWC
UDI-Device Identifier00815432022045
UDI-Public00815432022045
Combination Product (Y/N)N
PMA/510(K) Number
K151270
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1443-2501
Device Catalogue Number1443-2501
Device Lot Number101048
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2017
Initial Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-