CROSSROADS EXTREMITY SYSTEMS CROSSTIE; KEEL LOCK Back to Search Results
Model Number 7318-1818
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2020
Type of Reportable Event Malfunction
Event or Problem Description
At routine follow-up, a fractured keel-lock staple leg was identified in the follow-up x-ray.Tn fusion has occurred, and the patient is asymptomatic.The surgeon does not plan to revise the patient.
 
Additional Manufacturer Narrative
The lot history records of the sterile and individual component parts were reviewed in their entirety during the investigation.These records included quality inspection records, component inspection records, component print specification, component supplier certificates of conformance, supplier inspection certificates, raw material certificates of conformance, and supplier certificates of conformance.No non-conformances were found in any records that were reviewed that would have contributed to the device failure.7318-1818 500917.Keel-lock 18mm x 18 mm x 18mm sterile.240509-ep.103968.Keel-lock 18mm x 18 mm x 18mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSTIE
Common Device Name
KEEL LOCK
Manufacturer (Section D)
CROSSROADS EXTREMITY SYSTEMS
6423 shelby view dr.
suite 101
memphis TN 38134
Manufacturer Contact
6423 shelby view dr
suite 101
memphis, TN 38134
9012218406
MDR Report Key15501692
Report Number3020584246-2022-00048
Device Sequence Number7755397
Product Code HRS
UDI-Device Identifier00815432027736
UDI-Public00815432027736
Combination Product (Y/N)N
PMA/510(K) Number
K190722
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/04/2021
Device Model Number7318-1818
Device Catalogue Number7318-1818
Device Lot Number500917
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2020
Initial Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age43 YR
Patient SexFemale
-
-