EXACTECH, INC. OPTETRAK HI-FLEX TIBIAL INSERT SZ 4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
244-03-04 |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Unspecified Musculoskeletal problem (4535)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 4 years post op the initial right tka, this male patient was revised due to implant loosening bone loss and significant synovitis.Patient had been complaining of mechanical complications and chronic pain.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant device(s): 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t, 2921769; 200-02-35 - three peg patella 35mm, 4271300; 244-24-11 - optetrak hi-flex tibial insert sz 4 11mm, 1384194.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation results - the revision reported was likely the result of prosthesis wear and osteolysis.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.It is possible that a contributing factor to the reported wear may have been the result of being packaged in a non-conforming bag for more than 5 years before it was implanted.The extent and root cause of the prosthesis wear and osteolysis could not be determined as the devices were not returned for evaluation, and images and radiographs were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: a2, b4.The following sections were corrected: b1, b2, d1, d4, d8, h3, h4, h6 - health effect - clinical code, medical device problem code, component code, type of investigation, investigation findings and h11.The revision reported was likely the result of prosthesis wear and osteolysis.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.It is possible that a contributing factor to the reported wear may have been the result of being packaged in a non-conforming bag for more than 5 years before it was implanted.The extent and root cause of the prosthesis wear and osteolysis could not be determined as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
Per the right knee revision operative report, the patient was diagnosed with ¿osteolysis.¿ there was a massive amount of synovitis in the knee.Tissue was sent and this was negative for acute inflammation.The description of the operation noted that ¿the patellar component was well fixed.The femur and tibia were both well fixed but had a massive underlying osteolysis.¿ the operative note also states that there was ¿bony degeneration from poly wear.¿ the patient was taken to the recovery room having tolerated the procedure without complications.No further impact to the patient was reported.No additional information is available.
|
| |
|